Fit-Check

DeVilbiss Healthcare 525KS

seed_demo: impact history

Supply
187–253 V
Frequency
50 Hz
Draw
312 VA
Rated for
5–35 °C
Humidity
to 93%
Do not ship 1 blocker, 1 conditions

A screen, not a clearance. A person decides before anything ships.

Bardiya Primary Health Centre Nepal

230 V 50 Hz wards to 42.0°C 4 outages a week
1 blocked 1 conditional 5 unanswered 8 clear

What blocks it

Still missing, even if you solve that

Answer these and the screening sharpens

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.

Check Finding What would fix it Fact from Status
Operating temperature Rated to 35 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have. Bardiya Primary Health Centre would have to air-condition the room it lives in, or be sent a device rated for the heat. spec sheet FAIL
Grid reliability 4 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. a UPS sized to ride through an outage clinic profile MITIGABLE
Consumables Needs MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter. Whether these sell in Nepal is not known. Is there a supplier for MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter in Nepal? unknown NEEDS INFO
Spare parts Whether parts for DeVilbiss Healthcare 525KS can be sourced in Nepal is not known. Can spare parts for DeVilbiss Healthcare 525KS be bought in Nepal? unknown NEEDS INFO
Age and support The device's age is not known. What year was the device made? The nameplate usually prints it next to the serial number. unknown NEEDS INFO
Import and registration No import and registration rules for Nepal are on file. What are Nepal's import and registration rules for donated medical equipment? unknown NEEDS INFO
Shelf life on arrival No expiry date recorded for this consignment. Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? unknown NEEDS INFO
Supply frequency Rated 50 Hz, and the grid is 50 Hz. — spec sheet PASS
Supply voltage Rated 187-253 V, which covers the 230 V grid. — spec sheet PASS
Humidity Rated to 93 % and the site reaches 80 %. — spec sheet PASS
Medical oxygen The device does not need medical oxygen. — spec sheet PASS
Distilled water The device does not need distilled water. — spec sheet PASS
Compressed air The device does not need compressed air. — spec sheet PASS
Manual language The manual is available in English. — spec sheet PASS
Training and repair Simple enough to run with no biomedical technician on site. — spec sheet PASS

Saved as /verdict/e0ace929