Laerdal Medical LSU
seed_demo: impact history
- Supply
- 100–240 V
- Frequency
- 50/60 Hz
- Rated for
- 0–40 °C
- Humidity
- to 95%
Needs info
5 questions before a verdict
A screen, not a clearance. A person decides before anything ships.
Hopital Communautaire de Pignon Haiti
110 V 60 Hz wards to 36.0°C 12 outages a week
1 conditional
5 unanswered
9 clear
It would also need, once the questions are answered
- A voltage stabiliser
Answer these and the screening sharpens
- Is there a supplier for Serres suction bag 1000 ml (57 151), Serres canister 1000 ml (57 300), Serres suction tube CH25 180 cm (58 33 181) / disposable patient tubing 180 cm (78 04 12), Serres vacuum connector (78 12 06), Serres High Efficiency Filtration Kit (78 12 03), aerosol filter for reusable canister (78 12 00), float balls, vacuum seals, gaskets (packs of 10), LSU NiMH battery 12 V 2 Ah (78 08 00), replace after 3 years, AC power cord US/EU/UK, DC power cord 12-28 V in Haiti?
- Can spare parts for Laerdal Medical LSU be bought in Haiti?
- What year was the device made? The nameplate usually prints it next to the serial number.
- What are Haiti's import and registration rules for donated medical equipment?
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Grid reliability | 12 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | A voltage stabiliser | clinic profile | MITIGABLE |
| Consumables | Needs Serres suction bag 1000 ml (57 151), Serres canister 1000 ml (57 300), Serres suction tube CH25 180 cm (58 33 181) / disposable patient tubing 180 cm (78 04 12), Serres vacuum connector (78 12 06), Serres High Efficiency Filtration Kit (78 12 03), aerosol filter for reusable canister (78 12 00), float balls, vacuum seals, gaskets (packs of 10), LSU NiMH battery 12 V 2 Ah (78 08 00), replace after 3 years, AC power cord US/EU/UK, DC power cord 12-28 V. Whether these sell in Haiti is not known. | Is there a supplier for Serres suction bag 1000 ml (57 151), Serres canister 1000 ml (57 300), Serres suction tube CH25 180 cm (58 33 181) / disposable patient tubing 180 cm (78 04 12), Serres vacuum connector (78 12 06), Serres High Efficiency Filtration Kit (78 12 03), aerosol filter for reusable canister (78 12 00), float balls, vacuum seals, gaskets (packs of 10), LSU NiMH battery 12 V 2 Ah (78 08 00), replace after 3 years, AC power cord US/EU/UK, DC power cord 12-28 V in Haiti? | unknown | NEEDS INFO |
| Spare parts | Whether parts for Laerdal Medical LSU can be sourced in Haiti is not known. | Can spare parts for Laerdal Medical LSU be bought in Haiti? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Import and registration | No import and registration rules for Haiti are on file. | What are Haiti's import and registration rules for donated medical equipment? | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Supply frequency | Rated 50/60 Hz, and the grid is 60 Hz. | — | spec sheet | PASS |
| Supply voltage | Rated 100-240 V, which covers the 110 V grid. | — | spec sheet | PASS |
| Operating temperature | Rated to 40 C, and rooms reach 36 C. | — | spec sheet | PASS |
| Humidity | Rated to 95 % and the site reaches 90 %. | — | spec sheet | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | The device does not need distilled water. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in French. | — | spec sheet | PASS |
| Training and repair | Simple enough to run with no biomedical technician on site. | — | spec sheet | PASS |