Fit-Check

Laerdal Medical LSU

seed_demo: impact history

Supply
100–240 V
Frequency
50/60 Hz
Rated for
0–40 °C
Humidity
to 95%
Needs info 5 questions before a verdict

A screen, not a clearance. A person decides before anything ships.

Hopital Communautaire de Pignon Haiti

110 V 60 Hz wards to 36.0°C 12 outages a week
1 conditional 5 unanswered 9 clear

It would also need, once the questions are answered

Answer these and the screening sharpens

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.

Check Finding What would fix it Fact from Status
Grid reliability 12 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. A voltage stabiliser clinic profile MITIGABLE
Consumables Needs Serres suction bag 1000 ml (57 151), Serres canister 1000 ml (57 300), Serres suction tube CH25 180 cm (58 33 181) / disposable patient tubing 180 cm (78 04 12), Serres vacuum connector (78 12 06), Serres High Efficiency Filtration Kit (78 12 03), aerosol filter for reusable canister (78 12 00), float balls, vacuum seals, gaskets (packs of 10), LSU NiMH battery 12 V 2 Ah (78 08 00), replace after 3 years, AC power cord US/EU/UK, DC power cord 12-28 V. Whether these sell in Haiti is not known. Is there a supplier for Serres suction bag 1000 ml (57 151), Serres canister 1000 ml (57 300), Serres suction tube CH25 180 cm (58 33 181) / disposable patient tubing 180 cm (78 04 12), Serres vacuum connector (78 12 06), Serres High Efficiency Filtration Kit (78 12 03), aerosol filter for reusable canister (78 12 00), float balls, vacuum seals, gaskets (packs of 10), LSU NiMH battery 12 V 2 Ah (78 08 00), replace after 3 years, AC power cord US/EU/UK, DC power cord 12-28 V in Haiti? unknown NEEDS INFO
Spare parts Whether parts for Laerdal Medical LSU can be sourced in Haiti is not known. Can spare parts for Laerdal Medical LSU be bought in Haiti? unknown NEEDS INFO
Age and support The device's age is not known. What year was the device made? The nameplate usually prints it next to the serial number. unknown NEEDS INFO
Import and registration No import and registration rules for Haiti are on file. What are Haiti's import and registration rules for donated medical equipment? unknown NEEDS INFO
Shelf life on arrival No expiry date recorded for this consignment. Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? unknown NEEDS INFO
Supply frequency Rated 50/60 Hz, and the grid is 60 Hz. — spec sheet PASS
Supply voltage Rated 100-240 V, which covers the 110 V grid. — spec sheet PASS
Operating temperature Rated to 40 C, and rooms reach 36 C. — spec sheet PASS
Humidity Rated to 95 % and the site reaches 90 %. — spec sheet PASS
Medical oxygen The device does not need medical oxygen. — spec sheet PASS
Distilled water The device does not need distilled water. — spec sheet PASS
Compressed air The device does not need compressed air. — spec sheet PASS
Manual language The manual is available in French. — spec sheet PASS
Training and repair Simple enough to run with no biomedical technician on site. — spec sheet PASS

Saved as /verdict/cf0bd7c6