Edan Instruments SONOTRAX Basic A
seed_demo: impact history
- Rated for
- 5–40 °C
- Humidity
- to 80%
Do not ship
1 blocker, 1 conditions
A screen, not a clearance. A person decides before anything ships.
Bardiya Primary Health Centre Nepal
230 V 50 Hz wards to 42.0°C 4 outages a week
1 blocked
1 conditional
7 unanswered
6 clear
What blocks it
- Operating temperature. Rated to 40 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have.
Still missing, even if you solve that
- a UPS sized to ride through an outage
Answer these and the screening sharpens
- What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage.
- What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids.
- Is there a supplier for coupling gel (Edan P/N 11.57.14019), 2 x AA batteries (alkaline LR6 1.5 V or NiMH 1.2 V, can be bought locally) in Nepal?
- Can spare parts for Edan Instruments SONOTRAX Basic A be bought in Nepal?
- What year was the device made? The nameplate usually prints it next to the serial number.
- What are Nepal's import and registration rules for donated medical equipment?
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Operating temperature | Rated to 40 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have. | Bardiya Primary Health Centre would have to air-condition the room it lives in, or be sent a device rated for the heat. | spec sheet | FAIL |
| Grid reliability | 4 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | a UPS sized to ride through an outage | clinic profile | MITIGABLE |
| Supply frequency | The grid at Bardiya Primary Health Centre is 50 Hz. The device's rated frequency is not known. | What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage. | unknown | NEEDS INFO |
| Supply voltage | The grid at Bardiya Primary Health Centre is 230 V. The device's rated voltage range is not known. | What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids. | unknown | NEEDS INFO |
| Consumables | Needs coupling gel (Edan P/N 11.57.14019), 2 x AA batteries (alkaline LR6 1.5 V or NiMH 1.2 V, can be bought locally). Whether these sell in Nepal is not known. | Is there a supplier for coupling gel (Edan P/N 11.57.14019), 2 x AA batteries (alkaline LR6 1.5 V or NiMH 1.2 V, can be bought locally) in Nepal? | unknown | NEEDS INFO |
| Spare parts | Whether parts for Edan Instruments SONOTRAX Basic A can be sourced in Nepal is not known. | Can spare parts for Edan Instruments SONOTRAX Basic A be bought in Nepal? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Import and registration | No import and registration rules for Nepal are on file. | What are Nepal's import and registration rules for donated medical equipment? | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Humidity | Rated to 80 % and the site reaches 80 %. | — | spec sheet | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | The device does not need distilled water. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in English. | — | spec sheet | PASS |
| Training and repair | Simple enough to run with no biomedical technician on site. | — | spec sheet | PASS |