Draeger Medical Systems Isolette C2000
seed_demo: impact history
- Supply
- 108–132 V
- Frequency
- 50/60 Hz
- Draw
- 630 VA
- Rated for
- 20–30 °C
Do not ship
2 blockers, 2 conditions
A screen, not a clearance. A person decides before anything ships.
Bardiya Primary Health Centre Nepal
230 V 50 Hz wards to 42.0°C 4 outages a week
2 blocked
2 conditional
5 unanswered
6 clear
What blocks it
- Operating temperature. Rated to 30 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have.
- Training and repair. A high-skill device, and Bardiya Primary Health Centre has no biomedical technician on site.
Still missing, even if you solve that
- A step transformer at least 787 W, shipped with the device.
- a UPS sized to ride through an outage
Answer these and the screening sharpens
- Is there a supplier for distilled water for humidity reservoir (if humidity option fitted), air intake microfilter (change at least every 3 months), skin temperature probes, disposable (83 102 18 / 83 102 21), probe covers (Critter Covers), disposable access door / iris port cuffs, oxygen sensor cells (if oxygen option fitted; replace both together) in Nepal?
- Can spare parts for Draeger Medical Systems Isolette C2000 be bought in Nepal?
- What year was the device made? The nameplate usually prints it next to the serial number.
- What are Nepal's import and registration rules for donated medical equipment?
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Operating temperature | Rated to 30 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have. | Bardiya Primary Health Centre would have to air-condition the room it lives in, or be sent a device rated for the heat. | spec sheet | FAIL |
| Training and repair | A high-skill device, and Bardiya Primary Health Centre has no biomedical technician on site. | Bardiya Primary Health Centre would need operators trained on commissioning, and a service contract or a named regional technician, before this could run. | spec sheet | FAIL |
| Supply voltage | Rated 108-132 V, but the grid is 230 V. | A step transformer at least 787 W, shipped with the device. | spec sheet | MITIGABLE |
| Grid reliability | 4 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | a UPS sized to ride through an outage | clinic profile | MITIGABLE |
| Consumables | Needs distilled water for humidity reservoir (if humidity option fitted), air intake microfilter (change at least every 3 months), skin temperature probes, disposable (83 102 18 / 83 102 21), probe covers (Critter Covers), disposable access door / iris port cuffs, oxygen sensor cells (if oxygen option fitted; replace both together). Whether these sell in Nepal is not known. | Is there a supplier for distilled water for humidity reservoir (if humidity option fitted), air intake microfilter (change at least every 3 months), skin temperature probes, disposable (83 102 18 / 83 102 21), probe covers (Critter Covers), disposable access door / iris port cuffs, oxygen sensor cells (if oxygen option fitted; replace both together) in Nepal? | unknown | NEEDS INFO |
| Spare parts | Whether parts for Draeger Medical Systems Isolette C2000 can be sourced in Nepal is not known. | Can spare parts for Draeger Medical Systems Isolette C2000 be bought in Nepal? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Import and registration | No import and registration rules for Nepal are on file. | What are Nepal's import and registration rules for donated medical equipment? | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Supply frequency | Rated 50/60 Hz, and the grid is 50 Hz. | — | spec sheet | PASS |
| Humidity | Rated to 93 % and the site reaches 80 %. | — | IEC 60601-1 default | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | Needed, and Bardiya Primary Health Centre has it. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in English. | — | spec sheet | PASS |
Saved as /verdict/996a8ce9