Fit-Check

Draeger Medical Systems Isolette C2000

seed_demo: impact history

Supply
108–132 V
Frequency
50/60 Hz
Draw
630 VA
Rated for
20–30 °C
Do not ship 2 blockers, 2 conditions

A screen, not a clearance. A person decides before anything ships.

Bardiya Primary Health Centre Nepal

230 V 50 Hz wards to 42.0°C 4 outages a week
2 blocked 2 conditional 5 unanswered 6 clear

What blocks it

Still missing, even if you solve that

Answer these and the screening sharpens

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.

Check Finding What would fix it Fact from Status
Operating temperature Rated to 30 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have. Bardiya Primary Health Centre would have to air-condition the room it lives in, or be sent a device rated for the heat. spec sheet FAIL
Training and repair A high-skill device, and Bardiya Primary Health Centre has no biomedical technician on site. Bardiya Primary Health Centre would need operators trained on commissioning, and a service contract or a named regional technician, before this could run. spec sheet FAIL
Supply voltage Rated 108-132 V, but the grid is 230 V. A step transformer at least 787 W, shipped with the device. spec sheet MITIGABLE
Grid reliability 4 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. a UPS sized to ride through an outage clinic profile MITIGABLE
Consumables Needs distilled water for humidity reservoir (if humidity option fitted), air intake microfilter (change at least every 3 months), skin temperature probes, disposable (83 102 18 / 83 102 21), probe covers (Critter Covers), disposable access door / iris port cuffs, oxygen sensor cells (if oxygen option fitted; replace both together). Whether these sell in Nepal is not known. Is there a supplier for distilled water for humidity reservoir (if humidity option fitted), air intake microfilter (change at least every 3 months), skin temperature probes, disposable (83 102 18 / 83 102 21), probe covers (Critter Covers), disposable access door / iris port cuffs, oxygen sensor cells (if oxygen option fitted; replace both together) in Nepal? unknown NEEDS INFO
Spare parts Whether parts for Draeger Medical Systems Isolette C2000 can be sourced in Nepal is not known. Can spare parts for Draeger Medical Systems Isolette C2000 be bought in Nepal? unknown NEEDS INFO
Age and support The device's age is not known. What year was the device made? The nameplate usually prints it next to the serial number. unknown NEEDS INFO
Import and registration No import and registration rules for Nepal are on file. What are Nepal's import and registration rules for donated medical equipment? unknown NEEDS INFO
Shelf life on arrival No expiry date recorded for this consignment. Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? unknown NEEDS INFO
Supply frequency Rated 50/60 Hz, and the grid is 50 Hz. — spec sheet PASS
Humidity Rated to 93 % and the site reaches 80 %. — IEC 60601-1 default PASS
Medical oxygen The device does not need medical oxygen. — spec sheet PASS
Distilled water Needed, and Bardiya Primary Health Centre has it. — spec sheet PASS
Compressed air The device does not need compressed air. — spec sheet PASS
Manual language The manual is available in English. — spec sheet PASS

Saved as /verdict/996a8ce9