Philips Medical Systems SureSigns VM6
seed_demo: impact history
- Supply
- 100–240 V
- Frequency
- 50/60 Hz
- Draw
- 72 VA
- Rated for
- 10–40 °C
- Humidity
- to 80%
Do not ship
1 blocker, 1 conditions
A screen, not a clearance. A person decides before anything ships.
Bardiya Primary Health Centre Nepal
230 V 50 Hz wards to 42.0°C 4 outages a week
1 blocked
1 conditional
5 unanswered
8 clear
What blocks it
- Operating temperature. Rated to 40 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have.
Still missing, even if you solve that
- a UPS sized to ride through an outage
Answer these and the screening sharpens
- Is there a supplier for NBP cuffs and tubing, SpO2 sensors, ECG lead sets (3- and 5-lead), temperature probe, thermal recorder paper 58 mm (optional recorder), Li-ion smart battery 11.1 V 7200 mAh (989803144631), replace every 3 years / 300 cycles, NBP filter (4535 640 41171), fuses T1.6A 250V in Nepal?
- Can spare parts for Philips Medical Systems SureSigns VM6 be bought in Nepal?
- What year was the device made? The nameplate usually prints it next to the serial number.
- What are Nepal's import and registration rules for donated medical equipment?
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Operating temperature | Rated to 40 C, but rooms reach 42 C. It needs air conditioning the ward at Bardiya Primary Health Centre does not have. | Bardiya Primary Health Centre would have to air-condition the room it lives in, or be sent a device rated for the heat. | spec sheet | FAIL |
| Grid reliability | 4 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | a UPS sized to ride through an outage | clinic profile | MITIGABLE |
| Consumables | Needs NBP cuffs and tubing, SpO2 sensors, ECG lead sets (3- and 5-lead), temperature probe, thermal recorder paper 58 mm (optional recorder), Li-ion smart battery 11.1 V 7200 mAh (989803144631), replace every 3 years / 300 cycles, NBP filter (4535 640 41171), fuses T1.6A 250V. Whether these sell in Nepal is not known. | Is there a supplier for NBP cuffs and tubing, SpO2 sensors, ECG lead sets (3- and 5-lead), temperature probe, thermal recorder paper 58 mm (optional recorder), Li-ion smart battery 11.1 V 7200 mAh (989803144631), replace every 3 years / 300 cycles, NBP filter (4535 640 41171), fuses T1.6A 250V in Nepal? | unknown | NEEDS INFO |
| Spare parts | Whether parts for Philips Medical Systems SureSigns VM6 can be sourced in Nepal is not known. | Can spare parts for Philips Medical Systems SureSigns VM6 be bought in Nepal? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Import and registration | No import and registration rules for Nepal are on file. | What are Nepal's import and registration rules for donated medical equipment? | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Supply frequency | Rated 50/60 Hz, and the grid is 50 Hz. | — | spec sheet | PASS |
| Supply voltage | Rated 100-240 V, which covers the 230 V grid. | — | spec sheet | PASS |
| Humidity | Rated to 80 % and the site reaches 80 %. | — | spec sheet | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | The device does not need distilled water. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in English. | — | spec sheet | PASS |
| Training and repair | Simple enough to run with no biomedical technician on site. | — | spec sheet | PASS |