Roche Diabetes Care Accu-Chek Performa
seed_demo: impact history
- Rated for
- 8–44 °C
- Humidity
- to 90%
Needs info
6 questions before a verdict
A screen, not a clearance. A person decides before anything ships.
Kayunga District Hospital Uganda
240 V 50 Hz wards to 34.0°C 6 outages a week
1 conditional
6 unanswered
8 clear
It would also need, once the questions are answered
- A voltage stabiliser
Answer these and the screening sharpens
- What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage.
- What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids.
- Is there a supplier for Accu-Chek Performa test strips (store 2-30 °C, use by printed date, opened or unopened), Accu-Chek Performa control solution (discard 3 months after opening), lancets for lancing device, CR2032 3 V lithium coin cell in Uganda?
- Can spare parts for Roche Diabetes Care Accu-Chek Performa be bought in Uganda?
- What year was the device made? The nameplate usually prints it next to the serial number.
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Grid reliability | 6 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | A voltage stabiliser | clinic profile | MITIGABLE |
| Supply frequency | The grid at Kayunga District Hospital is 50 Hz. The device's rated frequency is not known. | What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage. | unknown | NEEDS INFO |
| Supply voltage | The grid at Kayunga District Hospital is 240 V. The device's rated voltage range is not known. | What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids. | unknown | NEEDS INFO |
| Consumables | Needs Accu-Chek Performa test strips (store 2-30 °C, use by printed date, opened or unopened), Accu-Chek Performa control solution (discard 3 months after opening), lancets for lancing device, CR2032 3 V lithium coin cell. Whether these sell in Uganda is not known. | Is there a supplier for Accu-Chek Performa test strips (store 2-30 °C, use by printed date, opened or unopened), Accu-Chek Performa control solution (discard 3 months after opening), lancets for lancing device, CR2032 3 V lithium coin cell in Uganda? | unknown | NEEDS INFO |
| Spare parts | Whether parts for Roche Diabetes Care Accu-Chek Performa can be sourced in Uganda is not known. | Can spare parts for Roche Diabetes Care Accu-Chek Performa be bought in Uganda? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Operating temperature | Rated to 44 C, and rooms reach 34 C. | — | spec sheet | PASS |
| Humidity | Rated to 90 % and the site reaches 85 %. | — | spec sheet | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | The device does not need distilled water. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in English. | — | spec sheet | PASS |
| Training and repair | Simple enough to run with nobody trained to run it and no biomedical technician on site. | — | spec sheet | PASS |
| Import and registration | Uganda's rules are on file. Labels: English (manual languages used as the label proxy). Paperwork to send: Prior no-objection from the Ministry of Health before shipment (item 11); Detailed equipment list: name and purpose, serial number, manufacturer, year of manufacture, last service certificate, quality assurance documents (CE, FDA, EU etc) and quantity (item 6); Detailed packing list: equipment name, quantity, serial number, weight and any special storage conditions (item 7). (MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, p.29-30) | — | spec sheet | PASS |