Fit-Check

Nellcor N-560

seed_demo: impact history

Supply
100–240 V
Draw
39 VA
Rated for
10–45 °C
Humidity
to 95%
Needs info 6 questions before a verdict

A screen, not a clearance. A person decides before anything ships.

Hopital Communautaire de Pignon Haiti

110 V 60 Hz wards to 36.0°C 12 outages a week
2 conditional 6 unanswered 7 clear

It would also need, once the questions are answered

Answer these and the screening sharpens

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.

Check Finding What would fix it Fact from Status
Grid reliability 12 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. A voltage stabiliser clinic profile MITIGABLE
Manual language The manual is in English; staff read French, Haitian Creole. Fit-Check generates a handover packet in French from the manufacturer's manual. A handover packet in French, built from the manufacturer's manual. spec sheet MITIGABLE
Supply frequency The grid at Hopital Communautaire de Pignon is 60 Hz. The device's rated frequency is not known. What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage. unknown NEEDS INFO
Consumables Needs Nellcor OXIMAX sensors only (single-patient adhesive MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I, MAX-R, MAX-FAST; OxiCliq A/N/P/I; Softcare SC-PR/SC-NEO/SC-A), reusable OXIMAX sensors (Durasensor DS-100A finger clip, Oxiband OXI-A/N, OXI-P/I), DOC-10 pulse oximetry cable, internal NiMH battery 9.6 V 3.8 Ah, fuses 2 A 250 V slow-blow 5x20 mm (qty 2). Whether these sell in Haiti is not known. Is there a supplier for Nellcor OXIMAX sensors only (single-patient adhesive MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I, MAX-R, MAX-FAST; OxiCliq A/N/P/I; Softcare SC-PR/SC-NEO/SC-A), reusable OXIMAX sensors (Durasensor DS-100A finger clip, Oxiband OXI-A/N, OXI-P/I), DOC-10 pulse oximetry cable, internal NiMH battery 9.6 V 3.8 Ah, fuses 2 A 250 V slow-blow 5x20 mm (qty 2) in Haiti? unknown NEEDS INFO
Spare parts Whether parts for Nellcor N-560 can be sourced in Haiti is not known. Can spare parts for Nellcor N-560 be bought in Haiti? unknown NEEDS INFO
Age and support The device's age is not known. What year was the device made? The nameplate usually prints it next to the serial number. unknown NEEDS INFO
Import and registration No import and registration rules for Haiti are on file. What are Haiti's import and registration rules for donated medical equipment? unknown NEEDS INFO
Shelf life on arrival No expiry date recorded for this consignment. Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? unknown NEEDS INFO
Supply voltage Rated 100-240 V, which covers the 110 V grid. — spec sheet PASS
Operating temperature Rated to 45 C, and rooms reach 36 C. — spec sheet PASS
Humidity Rated to 95 % and the site reaches 90 %. — spec sheet PASS
Medical oxygen The device does not need medical oxygen. — spec sheet PASS
Distilled water The device does not need distilled water. — spec sheet PASS
Compressed air The device does not need compressed air. — spec sheet PASS
Training and repair Simple enough to run with no biomedical technician on site. — spec sheet PASS

Saved as /verdict/34173649