Omron Healthcare HEM-907XL

seed_demo: impact history

Supply
120 V
Frequency
50/60 Hz
Draw
24 VA
Rated for
10–40 °C
Humidity
to 85%
Needs info 4 questions before a verdict

A screen, not a clearance. A person decides before anything ships.

Kayunga District Hospital Uganda

240 V 50 Hz wards to 34.0°C 6 outages a week
2 conditional 4 unanswered 9 clear

It would also need, once the questions are answered

Answer these and the screening sharpens

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.

Check Finding What would fix it Fact from Status
Supply voltage Rated 120-120 V, but the grid is 240 V. A step transformer at least 30 W, shipped with the device. spec sheet MITIGABLE
Grid reliability 6 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. A voltage stabiliser clinic profile MITIGABLE
Consumables Needs BP cuff/bladder sets S (17-22 cm), M (22-32 cm), L (32-42 cm), XL (42-50 cm), rechargeable battery pack HEM-907-PBAT (approx. 2-year life), air tube 1.3 m (HEM-TUBE-130XL), AC adapter HEM-ADPT907 (120 V). Whether these sell in Uganda is not known. Is there a supplier for BP cuff/bladder sets S (17-22 cm), M (22-32 cm), L (32-42 cm), XL (42-50 cm), rechargeable battery pack HEM-907-PBAT (approx. 2-year life), air tube 1.3 m (HEM-TUBE-130XL), AC adapter HEM-ADPT907 (120 V) in Uganda? unknown NEEDS INFO
Spare parts Whether parts for Omron Healthcare HEM-907XL can be sourced in Uganda is not known. Can spare parts for Omron Healthcare HEM-907XL be bought in Uganda? unknown NEEDS INFO
Age and support The device's age is not known. What year was the device made? The nameplate usually prints it next to the serial number. unknown NEEDS INFO
Shelf life on arrival No expiry date recorded for this consignment. Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? unknown NEEDS INFO
Supply frequency Rated 50/60 Hz, and the grid is 50 Hz. — spec sheet PASS
Operating temperature Rated to 40 C, and rooms reach 34 C. — spec sheet PASS
Humidity Rated to 85 % and the site reaches 85 %. — spec sheet PASS
Medical oxygen The device does not need medical oxygen. — spec sheet PASS
Distilled water The device does not need distilled water. — spec sheet PASS
Compressed air The device does not need compressed air. — spec sheet PASS
Manual language The manual is available in English. — spec sheet PASS
Training and repair Simple enough to run with nobody trained to run it and no biomedical technician on site. — spec sheet PASS
Import and registration Uganda's rules are on file. Labels: English (manual languages used as the label proxy). Paperwork to send: Prior no-objection from the Ministry of Health before shipment (item 11); Detailed equipment list: name and purpose, serial number, manufacturer, year of manufacture, last service certificate, quality assurance documents (CE, FDA, EU etc) and quantity (item 6); Detailed packing list: equipment name, quantity, serial number, weight and any special storage conditions (item 7). (MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, p.29-30) — spec sheet PASS

Saved as /verdict/2941e5f7