DeVilbiss Healthcare 525DS
seed_demo: impact history
- Supply
- 97–127 V
- Frequency
- 60 Hz
- Draw
- 310 VA
- Rated for
- 5–35 °C
- Humidity
- to 93%
Do not ship
1 blocker, 1 conditions
A screen, not a clearance. A person decides before anything ships.
Hopital Communautaire de Pignon Haiti
110 V 60 Hz wards to 36.0°C 12 outages a week
1 blocked
1 conditional
5 unanswered
8 clear
What blocks it
- Operating temperature. Rated to 35 C, but rooms reach 36 C. It needs air conditioning the ward at Hopital Communautaire de Pignon does not have.
Still missing, even if you solve that
- A voltage stabiliser
Answer these and the screening sharpens
- Is there a supplier for MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter in Haiti?
- Can spare parts for DeVilbiss Healthcare 525DS be bought in Haiti?
- What year was the device made? The nameplate usually prints it next to the serial number.
- What are Haiti's import and registration rules for donated medical equipment?
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest?
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag on the right says where the fact came from, which is the only part a model touched.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Operating temperature | Rated to 35 C, but rooms reach 36 C. It needs air conditioning the ward at Hopital Communautaire de Pignon does not have. | Hopital Communautaire de Pignon would have to air-condition the room it lives in, or be sent a device rated for the heat. | spec sheet | FAIL |
| Grid reliability | 12 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | A voltage stabiliser | clinic profile | MITIGABLE |
| Consumables | Needs MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter. Whether these sell in Haiti is not known. | Is there a supplier for MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter in Haiti? | unknown | NEEDS INFO |
| Spare parts | Whether parts for DeVilbiss Healthcare 525DS can be sourced in Haiti is not known. | Can spare parts for DeVilbiss Healthcare 525DS be bought in Haiti? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Import and registration | No import and registration rules for Haiti are on file. | What are Haiti's import and registration rules for donated medical equipment? | unknown | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Supply frequency | Rated 60 Hz, and the grid is 60 Hz. | — | spec sheet | PASS |
| Supply voltage | Rated 97-127 V, which covers the 110 V grid. | — | spec sheet | PASS |
| Humidity | Rated to 93 % and the site reaches 90 %. | — | spec sheet | PASS |
| Medical oxygen | The device does not need medical oxygen. | — | spec sheet | PASS |
| Distilled water | The device does not need distilled water. | — | spec sheet | PASS |
| Compressed air | The device does not need compressed air. | — | spec sheet | PASS |
| Manual language | The manual is available in French. | — | spec sheet | PASS |
| Training and repair | Simple enough to run with no biomedical technician on site. | — | spec sheet | PASS |
Saved as /verdict/191e8c69