HackTN 2026 prototype with demo data. Nothing is shipped and no one is emailed.

Fit-Check
Demo: switch viewpoint. There are no logins.

← 1 oxygen concentrator for Kayunga District Hospital

DeVilbiss Healthcare 525KS

oxygen concentrator · 1 available from Midland Teaching Hospital, surplus sample, Leeds, United Kingdom

6694 km · 4159 mi to Kayunga District Hospital, about 61 days to deliver

Supply
187–253 V
Frequency
50 Hz
Draw
312 VA
Rated for
5–35 °C
Humidity
to 93%
Made
2019
Ship with 2 things it cannot arrive without

A screen, not a clearance. A person decides before anything ships.

Kayunga District Hospital Uganda sample

240 V 50 Hz wards to 34.0°C 6 outages a week
2 conditional 13 clear
Generate handover packet

This is seeded demo data and stays as it is. Register your own clinic or post your own offer to try this.

Must go with it

A voltage stabilizer covered: 4 available

covers Sollatek AVS30 (Kampala Power Trust), 4 available — its own plate runs on 240 V 50 Hz

won't work here itself automatic voltage regulator (Houston surplus AVR) — its own plate fails 240 V 50 Hz; it does not count

2 years of MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter. no donor has one yet

Audit trail — 15 checks

Every row is decided by code, not by a model. The tag says where the fact came from, and the link under it opens the page it was read on.

Check Finding What would fix it Fact from Status
Grid reliability 6 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. A voltage stabilizer clinic profile MITIGABLE
Consumables Needs MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter, which are not sold in Uganda. The device works until the first box runs out, then stops. 2 years of MC44D-605 intake bacteria filter, 303DZ-605 cabinet air filter, PV5LD-651 final bacteria filter, 525DD-626 compressor filter, shipped with the device, and a named supplier for after that. recorded by the nonprofit spec sheet EC Declaration of Conformity 525KS/525KS-LT/525KS-UK/525PS (2024), p.1the four filters of the 525 series; sieve bed 525D-619 and compressor rebuild kit 525D-643 are spare parts, not consumables (see notes) recorded by Fit-Check field team, 2026-11-10 MITIGABLE
Supply frequency Rated 50 Hz, and the grid is 50 Hz. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.11SE-525KS-2 Rev A (2022) also lists a 230 V / 60 Hz 525KS build; the 2025 editions list 50 Hz only. Check the unit's rating label PASS
Supply voltage Rated 187-253 V, which covers the 240 V grid. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.11 PASS
Operating temperature Rated to 35 C, and rooms reach 34 C. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.11 PASS
Humidity Rated to 93 % and the site reaches 85 %. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.11 PASS
Medical oxygen The device does not need medical oxygen. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-2 Rev A (2022), p.6the concentrator makes its oxygen from room air with its own compressor; the manual does not say in so many words that no piped oxygen is needed PASS
Distilled water The device does not need distilled water. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-2 Rev A (2022), p.8no water requirement in the owner's manual; the bubble humidifier is an optional accessory PASS
Compressed air The device does not need compressed air. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-2 Rev A (2022), p.6the concentrator makes its oxygen from room air with its own compressor; the manual does not say in so many words that no piped compressed air is needed PASS
Spare parts Parts can be sourced in Uganda. — recorded by the nonprofit recorded by Fit-Check field team, 2026-11-10 PASS
Age and support Made in 2019, so it is 7 years old with working life left. — stated by the donor PASS
Manual language The manual is available in English. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.221 languages across three editions: SE-525KS-1 (en es fr de it nl tr pt pl ar), SE-525KS-2 (en el da sv fi no cs sk zh), SE-525KS-3 (en sl ro ms) PASS
Training and repair Simple enough to run with nobody trained to run it and no biomedical technician on site. — spec sheet DeVilbiss 525KS Instruction Guide SE-525KS-2 Rev A (2022), p.6author's reading of the intended use: supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities; the manual does not grade operator skill PASS
Import and registration Uganda's rules are on file. Labels: English (manual languages used as the label proxy). Paperwork to send: Prior no-objection from the Ministry of Health before shipment (item 11); Detailed equipment list: name and purpose, serial number, manufacturer, year of manufacture, last service certificate, quality assurance documents (CE, FDA, EU etc) and quantity (item 6); Detailed packing list: equipment name, quantity, serial number, weight and any special storage conditions (item 7). (MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, p.29-30) — spec sheet import rule MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, pp.29-30 DeVilbiss 525KS Instruction Guide SE-525KS-1 Rev D (2025), p.221 languages across three editions: SE-525KS-1 (en es fr de it nl tr pt pl ar), SE-525KS-2 (en el da sv fi no cs sk zh), SE-525KS-3 (en sl ro ms) PASS