← 1 oxygen concentrator for Kayunga District Hospital
Unnamed oxygen concentrator
oxygen concentrator · 2 available from Bayou community clinic network sample, Houston, United States
13577 km · 8436 mi to Kayunga District Hospital, about 70 days to deliver
A screen, not a clearance. A person decides before anything ships.
Kayunga District Hospital Uganda sample
240 V 50 Hz wards to 34.0°C 6 outages a weekWould go with it, once the questions are answered
covers Sollatek AVS30 (Kampala Power Trust), 4 available — its own plate runs on 240 V 50 Hz
won't work here itself automatic voltage regulator (Houston surplus AVR) — its own plate fails 240 V 50 Hz; it does not count
11 open questions before a verdict
- What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage. (for the donor)
- What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids. (for the donor)
- Does the device need a medical oxygen supply to operate? (for the donor)
- Does the device need distilled or deionised water? (for the donor)
- Does the device need a medical compressed air supply? (for the donor)
- What does the device consume in use - reagents, cartridges, circuits, electrodes, filters? (for the donor)
- Can spare parts for this device be bought in Uganda? Record the nonprofit's answer
- What year was the device made? The nameplate usually prints it next to the serial number. (for the donor)
- What languages is the manual available in, and is one included? (for the donor)
- Does running this device need specialist training, and does servicing it need a biomedical technician? (for the donor)
- Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? (for the donor)
Audit trail — 15 checks
Every row is decided by code, not by a model. The tag says where the fact came from, and the link under it opens the page it was read on.
| Check | Finding | What would fix it | Fact from | Status |
|---|---|---|---|---|
| Grid reliability | 6 outages a week. Switching transients kill power supplies in equipment that assumes a steady feed. | A voltage stabilizer | clinic profile | MITIGABLE |
| Supply frequency | The grid at Kayunga District Hospital is 50 Hz. The device's rated frequency is not known. | What supply frequency is the device rated for? The nameplate prints it in Hz, next to the voltage. | unknown | NEEDS INFO |
| Supply voltage | The grid at Kayunga District Hospital is 240 V. The device's rated voltage range is not known. | What voltage range is the device rated for? A nameplate reading 100-240 V means it handles both grids. | unknown | NEEDS INFO |
| Medical oxygen | Not known whether the device needs medical oxygen. | Does the device need a medical oxygen supply to operate? | unknown | NEEDS INFO |
| Distilled water | Not known whether the device needs distilled water. | Does the device need distilled or deionised water? | unknown | NEEDS INFO |
| Compressed air | Not known whether the device needs compressed air. | Does the device need a medical compressed air supply? | unknown | NEEDS INFO |
| Consumables | Not known what this device consumes in normal use. | What does the device consume in use - reagents, cartridges, circuits, electrodes, filters? | unknown | NEEDS INFO |
| Spare parts | Whether parts for this device can be sourced in Uganda is not known. | Can spare parts for this device be bought in Uganda? | unknown | NEEDS INFO |
| Age and support | The device's age is not known. | What year was the device made? The nameplate usually prints it next to the serial number. | unknown | NEEDS INFO |
| Manual language | Staff at Kayunga District Hospital read English. No manual could be found and the manual's language is not known. | What languages is the manual available in, and is one included? | unknown | NEEDS INFO |
| Training and repair | How much skill the device needs is not known. | Does running this device need specialist training, and does servicing it need a biomedical technician? | unknown | NEEDS INFO |
| Import and registration | Uganda requires labels in English, and the manual's language is not known (manual languages used as the label proxy). (MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, p.29-30) | What languages is the manual available in, and is one included? | unknown import rule MoH Uganda, National Medical Equipment Guidelines, November 2024, section 3.6 Donated Equipment and Supplies, pp.29-30 | NEEDS INFO |
| Shelf life on arrival | No expiry date recorded for this consignment. | Does anything being sent carry an expiry date - sterile circuits, filters, reagents, batteries? If so, what is the earliest? | unknown | NEEDS INFO |
| Operating temperature | Rated to 40 C, and rooms reach 34 C. | — | IEC 60601-1 default | PASS |
| Humidity | Rated to 93 % and the site reaches 85 %. | — | IEC 60601-1 default | PASS |